Skip to content
CT Tube Replacement for Siemens CT Scanner (Somatom go.Now) TrueRay Chorus
Quotation
CT Tube Replacement for Siemens(Somatom go. Now)
- Work environment:0~ 40 ºC,30% ~75 %,700 hPa ~ 1,060 hPa
- Oil running cooling:
Temperature:10-50℃
Nominal Voltage:140kV
Over voltage:154kV
Cathode emission:
Voltage: 130kV
Filament current: 5.6A
- Tube current:
Small focus F1: ≥60mA
Large focus F2: ≥70mA
- Filament character:
Filament current: 0-5.6A
- Filament Voltage:
Small focus F1: 0-6.6V
Large focus F2: 0- 11.4V
- Nominal anode input power:
Small focus F1:28kW
Large focus F2 :34kW
Average power: 2.5kW
- Nominal focal spot size:
Small focus F1:0.8×0.4mm
Large focus F2:0.8×0 7mm
Output dose: ≥3.5mGy/s
Permanent filtration: 0.5mmTi±0. 15mmTi
Leakage current to ground: ≤2mA
Anode rotation speed: ≥8,400r/min
Anode capacity: 3.5 MHU
Anode target diameter: 140 mm ±0.5mm
Anode angle: 7.0°±0.5°
Weight: 8.2 kg±0. 1kg
Reliability: 200,000ss