MRI Compatible Patient Monitoring with ECG UZZMR-II

Quotation

1.Host
Function: ECG, non-invasive blood pressure and oxygen saturation monitoring, alarm and settings
Display: Color TFT capacitive touch screen, screen size: 12 inches, 1024×768, LED

2. ECG
Type CF (defibrillation proof)
Signal acquisition: ECG sensor, electrode loss detection
Patient leakage power < 10 uA
Heart rate measurement range: 30-250bpm, ±2bpm
Accuracy: ± 10% or ± 5 BPM
Lead configuration: I, II, III, V, AVF, AVR, AVL

3. Alarm
Instructions: Sound and Vision
Gears: High, Medium, Low and Information messaging
Volume: Adjustable, 4 levels of loudness adjustable, or off
Mute: permanent or timed for 2 minutes

4.Safety
Defibrillator Protection: Up to 5 KV, 15 seconds recovery

5.Power supply
Voltage range: 85 – 264 VAC
Host battery capacity: 14.8V, 6 Ah lithium polymer

6.Working environment
Temperature range: +10℃ to +40℃
Humidity range: 5% to 85% RH, non-condensing
Altitude range: sea level to 5000 meters

7. Blood oxygen saturation
Saturation range: 1% – 100%
Saturation accuracy: ±3% (70%-100%), ±1% (<70%)
Pulse rate range: 30 – 240 ppm
Pulse rate accuracy: ±3 ppm
Pulse rate resolution: 1 PPM
Wavelength range: Red: 660nm IR: 900nm
Calibration range: 70 – 100%
8.NIBP non-invasive blood pressure
Mode: manual, automatic
Measurement time: < 60 seconds typical; standard adult cuff, stretch rate approximately 4 mmHg/second, in steps.
Systolic blood pressure:
Adult: 40 – 260 mmHg
Children: 40 – 230 mmHg
Newborn: 40 – 130 mmHg
MAP: Adult: 26 – 220 mmHg
Children: 26 – 183 mmHg
Newborn: 26 – 110 mmHg
Diastolic blood pressure:
Adult: 20 – 200 mmHg
Children: 20 – 160 mmHg
Newborn: 20 – 100 mmHg
Cuff pressure measurement range: 0 to 300 mmHg
Pulse rate range: 30 to 220BPM
Pulse rate accuracy: ±3 BPM
9.Physical dimensions
Length x width x height: 520mm*460mm*1400mm
Weight: 23kg