Transcranial Magnetic Stimulator XY-K-JLC-D

SKU: XY-K-JLC-D Category:

Technical Parameter

1. All-in-one host, the liquid cooling system is highly integrated with the stimulation host
2. Rated input power: 2500VA
3. Performance:
3.1. The maximum magnetic field output intensity is 3T, the tolerance is ±20%; the stimulation intensity is adjustable from 0 to 100%, and the step is 1%
3.2. The output pulse frequency is 0Hz~100Hz, when 0Hz~1Hz, the step is 0.1Hz; when 1Hz~100Hz, the step is 1Hz, the actual output frequency tolerance is ±10%
3.3. The output pulse width should be in the range of 100~200μs
3.4. The stimulation mode is divided into manual mode and automatic program mode
Manual mode: including single stimulation and continuous stimulation, triggered by a single stimulation on the lower computer or PC or the start button on the PC;
Automatic program mode: select the treatment prescription in the plan list on the PC, and trigger it by pressing the start button.
3.5. Time parameters:
a) Number of strings: 1-250
b) Interval time: 0-60s, step 1s
c) Cycle group number: 1-250

Shopping Cart