150kV Rotating Anode X‐Ray Tube Assembly MWHX7620A

SKU: MWHX7620A Category:

Technical Parameter

  1. X-ray tube assembly for all routine diagnostic examinations with conventional or digital radiographic and fluoroscopic workstations
  2. The insert features: 12° Rhenium-Tungsten molybdenum target (RTM)
  3. Focal spots: Small 0.6, Large: 1.2
  4. Maximum tube voltage: 150kV
  5. Accommodated with IEC60526-type high-voltage cable receptacles
  6. High voltage generator should accord with IEC60601-2-7
  7. IEC Classification (IEC 60601-1:2005): Class I ME EQUIPMENT 
    ●The designation is composed as follows:
    MWHX7620TubeAHigh voltage socket with 90 degree direction
    MWTX76-0.6/1.2-150BHigh voltage socket with 270 degree direction
    Technical Data
    PropertySpecificationStandard
    Nominal input power(s) of the anodeF1F2IEC 60613
    20kW(50/60Hz)50kW(50/60Hz) -
    Anode heat storage capacity210kJ (300 kHU)IEC 60613
    Maximum cooling capacity of the anode870W -
    Heat storage capacity900kJ -
    Max. continuous heat dissipation without Air-circular180W -
    Anode material
    Anode top coating material
    Rhenium-Tungsten-TZM(RTM)
    Rhenium-Tungsten-(RT)
    -
    Target angle (Ref: reference axis)12°IEC 60788
    X-ray tube assembly inherent
    filtration
    1.5 mm Al/75kVIEC 60601-1-3
    Focal spot nominal value(s)F1(small focus)F2(large focus)IEC 60336
    0.61.2 -
    X-ray tube nominal voltage
    Radiographic
    Fluoroscopic
    150kV
    125kV
    IEC 60613
    Data on cathode heating≈/AC,<20 kHz -
    Max.currentF1F2 -
    Max voltage5.2A
    ≈6.8~9.2V
    5.4A
    ≈11.9~16.1V
    -
    Leakage radiation at 150 kV / 3mA in
    1m distance
    ≤1.0 mGy/hIEC60601-1-3
    Maximum radiation field430X430mm at SID 1m -
    X-ray tube assembly weightApprox. 18 kg -