Technical Parameter
- X-ray tube assembly for all routine diagnostic examinations with conventional or digital radiographic and fluoroscopic workstations
- The insert features: 12.5° Rhenium-Tungsten molybdenum target (RTM)
- Focal spots: Small 0.6, Large: 1.2
- Maximum tube voltage: 125kV
- Accommodated with IEC60526 type high-voltage cable receptacles
- High voltage generator should accord with IEC60601-2-7
- IEC Classification (IEC 60601-1:2005): Class I ME EQUIPMENT
Property Specification Standard Nominal input power(s) of the anode F1
20kW(50/60Hz)F2
40kW(50/60Hz)IEC 60613 Anode heat storage capacity 110 kJ (150 kHU) 110 kJ (150 kHU) IEC 60613 Maximum cooling capacity of the anode 500W 500W Heat storage capacity 900kJ 900kJ Max.continuous heat dissipation without Air-circular 180W 180W Anode material
Anode top coating materialRhenium-Tungsten-TZM(RTM)
Rhenium-Tungsten-(RT)Rhenium-Tungsten-TZM(RTM)
Rhenium-Tungsten-(RT)Target angle (Ref: reference axis) 12.5° 12.5° IEC 60788 X-ray tube assembly inherent filtration 1.5 mm Al / 75kV 1.5 mm Al / 75kV IEC 60601-1-3 Focal spot nominal value(s) F1(small focus)
0.6F2(large focus)
1.2IEC 60336 X-ray tube nominal voltage
Radiographic
Fluoroscopic125kV
100kV125kV
100kVIEC 60613 Data on cathode heating
≈/AC,< 20 kHz ≈/AC,< 20 kHz F1 F2 Max.current
Max voltage5.1A
≈7~9V5.1A
≈12~14VLeakage radiation at 150 kV / 3mA in
1m distance≤1.0 mGy/h ≤1.0 mGy/h IEC60601-1-3 Maximum radiation field 443X443mm at SID 1m 443X443mm at SID 1m X-ray tube assembly weight Approx. 18 kg Approx. 18 kg