Nominal input power(s) of the anode | F1
21kW(50/60Hz) | F2
42.5kW(50/60Hz) | IEC 60613 |
Anode heat storage capacity | 100 kJ (140kHU) | 100 kJ (140kHU) | IEC 60613 |
Maximum cooling capacity of the anode | 475W | 475W | |
Heat storage capacity | 900kJ | 900kJ | |
Max. continuous heat dissipation without Air-circular | 180W | 180W | |
Anode material
Anode top coating material | Rhenium-Tungsten-TZM(RTM)
Rhenium-Tungsten-(RT) | Rhenium-Tungsten-TZM(RTM)
Rhenium-Tungsten-(RT) | |
Target angle (Ref: reference axis) | 16° | 16° | IEC 60788 |
X-ray tube assembly inherent
filtration | 1.5 mm Al / 75kV | 1.5 mm Al / 75kV | IEC 60601-1-3 |
Focal spot nominal value(s) | F1(small focus)
1.0 | F2(large focus)
2.0 | IEC 60336 |
X-ray tube nominal voltage
Radiographic
Fluoroscopic | 125kV
100kV | 125kV
100kV | IEC 60613 |
Data on cathode heating
Max.current
Max voltage | ≈/AC,< 20 kHz | ≈/AC,< 20 kHz | |
| F1 | F2 | |
Max.current
Max voltage | 5.1A
≈5.8~7.8V | 5.1A
≈7.7~10.4V | |
Leakage radiation at 150 kV / 3mA in
1m distance | ≤1.0mGy/h | ≤1.0mGy/h | IEC60601-1-3 |
Maximum radiation field | 573x573mm at SID 1m | 573x573mm at SID 1m | |
X-ray tube assembly weight | Approx. 18 kg | Approx. 18 kg | |